INTRODUCTION

Gut-on-a-chip models offer a powerful in vitro platform for studying the physiology and patho-physiology of the intestine. The gut is home to a microbiome that plays an important role in health and disease. Some of these microorganisms survive the hypoxic conditions found in the intestine. Most commensal microbes in the digestive system are anaerobic and require low levels of oxygen1. The BE-FLOW microfluidic device is made of a Cyclic Olefin Polymer (COP). This material is impermeable to gases and allows the control of gas concentration within the devices. Overall, this system offers a way to control the microenvironment of the intestinal epithelium that is closer to the physiological state.

The peristaltic movement of the intestinal wall is known to affect the differentiation of the cells that line the digestive tract2. It can also impact the infectivity of human pathogens such as Shi- gella3. The local forces exerted onto the intestinal epithelium range from 0.02-0.08 dyn/cm2. It is important to apply similar forces in a controlled manner when studying intestinal physiology and pathophysiology. Devices that generate fluid flow and shear stress have been reported to promote accelerated differentiation of intestinal epithelial cells, formation of three-dimensional villus-like structures, and enhanced intestinal barrier function, mimicking the complex functions of the normal human intestine 4. By combining BE-FLOW chip with Fluigent pressure-based flow controllers, we propose a complete system to set up a model of human gut-on-chip.

MATERIALS

1. BEONCHIP device: BE-Flow

The design of the BE-FLOW consists of 2 independent channels with threaded inlets and outlets enabling the insertion of connectors and tubing coupled to fluidic controllers. It is possible to apply an independent flow rate in both channels using any perfusion system. To avoid culture media evaporation during the onset of the experiment (before perfusion has started), water-filled reservoirs are positioned next to the medium reservoir (cf scheme).

 2.FLUIGENT equipment

Fluigent positive pressure controllers precisely regulate the flow rate of a microfluidic system by applying small air pressures (mbar) over a closed reservoir to induce the rise of a solution through an output tube.

  • Air pressure source. This can be a compressed air intake or external compressor. The use of the external pressure compressor (FLPG or RX) will facilitate the assembly of the system under a biosafety cabinet to perform cell culture experiments under sterile conditions. Once the system is assembled and placed inside the incubator, it can be connected to a fixed compressed air intake.

  • Flow-EZ pressure controller. The pressure controller is capable of regulating the air inlet to the liquid reservoir to displace it.

  • Link / Power supply module.The link module (linked to the FlowEZ pressure controller) allows the connection to a PC for software control.

  • CAPs. Caps for the hermetic closure of reservoirs.

  • FlowUnit. Linking a flow sensor to the pressure controller allows direct control over flow rate. The flow sensor will always be placed after the outlet reservoir and before the microfluidic device.

 The fluidic system configuration is represented in the illustration:

Pressure compressor RX
Regulator
FlowEZ (1000 mbar)
FlowUnit S

During assembly, connect the Flow Unit in the direction of flow, indicated by the arrow on the Flow Unit. Before starting the experiment, calibrate the pressure to ‘0’ on the pressure controller. To do this, manually set a pressure of ‘0’ and click on the option “pressure calibrate” from the menu in the controllers.

Use the P=0 button

METHOD AND PROTOCOL

1. BE-FLOW cell culture
Before cell seeding, pre-warm the BE-FLOW standard device in the incubator overnight to
minimize the formation of air bubbles.

  1. Fill the channel with 100μl of collagen at 0.1 mg/ml (in PBS 1X) and incubate at 37°C for 30 minutes. Wash the channel by adding 100 μl of PBS 1X into the inlet and by aspirating it at the outlet using P200 pipette. Repeat this step three times.
  2. Aspirate PBS 1X completely before cell seeding.
  3. Seed 1.106 Caco-2 cells resuspended in 50μl of culture media.
  4. Incubate at 37°C, 5% CO2 until the cells have properly spread (2-4h).
  5. Once cell spreading has occurred, add 300μl of culture medium into the medium reservoirs. Add H2O to the evaporation reservoirs, cover and keep it in the incubator until the perfusion set up is ready to be connected (around 2h).

If connection to a perfusion system is not possible for more than 2 hours following cell seeding/cell spreading, keep the device in the incubator and renew culture medium using a rocker as long as needed. Check the channel has enough fluid and does not dry up.

2. Microfluidic set up

Before setting the flow up:

  • Sterilize and pre-warm the tubing and the fluidic elements overnight at 37°C.
  • Put the system under a laminar flow cabinet.
  • The channels and inlets/outlets should never be depleted of culture medium.
  • Both inlets and outlets use connectors (1/4’’ – 28).
    To work with cells, the assembly of the circuit must be carried out under sterile conditions
    under a biosafety cabinet. Before use, autoclave the CAPs, pneumatic tubes, threaded connections and ferrules to be used.
  1. Connect all circuit components except the microfluidic device.
  2. Establish a flow of ethanol (70%) for at least 15 min to sterilize the sensor and then wash with abundant sterile H2O. Dry by passing air through at maximum pressure. The sensor CANNOT be autoclaved.
  3. Prime the tubes of the circuit with culture medium until there are no air bubbles.
  4. Remove the culture medium from the chip reservoirs (not from inlet/outlet wells) and connect the tubes to the outlet/inlet using the threaded connections and ferrules.
    The ferrules should be manipulated with the help of sterile forceps.
    Ensure that the tube is perfectly fixed. Remove any displaced medium from the reservoirs.
  5. Once the system is closed, set the flow to 5.8 μl/min (shear stress 0.02 dyn/cm2) for 4 days to promote villi formation. Observe the system under perfusion for a few minutes to check that there are no leaks.

For more information check the following tutorials:
Microfluidic set-up

Flow EZ: Getting started

Flow EZ: Insert in my setup

Flow EZ: Add my FLOW UNIT

Connecting a microfluidic control system


RESULTS

After 4 days of perfusion, flow is stopped and the system is placed under an inverted microscope to monitor cell behaviour/differentiation. As observed in Figure 1, cell density is increased after 4 days of culture under flow conditions. Furthermore, Caco-2 started to form 3D structures, features not seen under static conditions.

AUTHORS

This article has been written by Sandra González Lana.

BIBLIOGRAPHY

1   N. Ashammakhi, R. Nasiri, N. R. de Barros, P. Tebon, J. Thakor, M. Goudie, A. Shamloo, M. G. Martin and A. Khademhosseni, Biomaterials, 2020. DOI:10.1016/j.biomaterials.2020.120196.
2. H. J. Kim, D. Huh, G. Hamilton and D. E. Ingber, Lab Chip, 2012. DOI: 10.1039/c2lc40074j

3. A. Grassart, V. Malardé, S. Gobaa, A. Sartori-Rupp, J. Kerns, K. Karalis, B. Marteyn, P. San

sonettti, N. Sauvonnet. Cell Cell Host Microbe. 2019. DOI: 10.1016/j.chom.2019.08.007

4. HJ. Kim, DE. Ingber, Integr Biol (Camb), 2013. DOI: 10.1039/c3ib40126j5.

Cell fixation, staining and immunostaining in a chip.

Introduction Cell fixation and immunostaining are critical techniques in cell biology and biomedical research. Cell fixation is the process of immobilising cells in a particular state, preserving their morphology, and preventing any further changes. This is crucial...

Organoids on chip

What is an organoid and why use it in research? There has been an increasing shift towards the development of 3D cell culture models in attempts to create an increased complexity that can be compared with the in vivo better than 2D models. From the different 3D...

BE-Transflow device as an epithelium-on-a-chip model for permeability studies

Researchers from the University of Zaragoza have employed the Be-Transflow device to develop a stratified epithelia-on-a-chip model, highlighting this device as a useful platform for permeability studies.     Replicating the permeability of a tissue barrier is pivotal...

BE-Gradient Barrier-Free applications: Hydrogel confinement and diffusion profile

Introduction Our new design BE-Gradient Barrier-Free is a device designed for 3D culture where a central chamber is linked with two fluidic lateral channels (Figure 1). The innovation we present is the absence of any physical barrier between the central chamber and...

BE-Doubleflow App. notes: Gut-on-chip 2

Introduction In our previous technical note “BE-Doubleflow App. notes: Gut-on-chip 1” we explored our device BE-Doubleflow for gut-on-a-chip (GOC) models in collaboration with AINIA. In that note we compared our device with an insert platform with GOC outperforming...